KAYDEN® CLASSIC® Thermal Flow, Level, Interface Switch, Modbus, 1/2" NPT, 150 cm U Length, 316 SST Sensor, Local Electronics Enclosure with Blind Cover, Approvals: CSA

Price
$490.40 USD
Stock
25+ In Stock

  • Sensor Material - 316/316L Stainless Steel
  • Process Connection - Male NPT - 1/2"
  • Insertion 'U' Length - 150 Custom 'U' Length: Centimeters
  • Cover for Local Enclosure - Blind Cover
  • Remote Electronics Enclosure & Cover - Not Required
  • Agency Approvals - CSA (Canada & US)

Specifications

Accuracy

  • Flow Service: ± 1.0% of Set Point Velocity over Operating Range of ± 28° C (± 50° F)
  • Level Service: ± 0.64 cm (0.25")

Applications

  • Flow
  • Interface
  • Level

Approvals & Certifications

  • CRN - 316, 316L Stainless Steel - Canadian Registration Number 0F20731.5C: 2053 psi, 14,155 kPa, 142 bar
  • CRN - Hastelloy® C-276 - Canadian Registration Number 0F20731.5C: 2290 psi, 15,790 kPa, 158 bar
  • CRN - Titanium - Canadian Registration Number 0F20731.5C: 1165 psi, 8033 kPa, 81 bar
  • CSA - Local (Sensor) Enclosure Flameproof: Class I, Division 1, Groups B, C, D, Ex D IIB + H2 AEx D IIB + H2 (Class I, Zone 1, Group IIB + H2) Temperature Code: T3 Enclosure Typical 4, IP55 Single Seal Approved Per ANSI/ISA 12.27.01-2003
  • CSA - Remote Electronics Enclosure - Optional Flameproof: Class I, Groups B, C, D, Ex D IIB + H2 AEx D IIB + H2 (Class I, Zone 1, Groups IIB + H2) Temperature Code: T3
  • FM - Local (Sensor) Enclosure Flameproof: Class I, Division 1, Groups B, C, D Temperature Codes: T2D (Ta=75° C), T3 (Ta=65° C) Class I, Zone 1, AEx D IIB + H2 Enclosure Typical 4/IP55
  • FM - Remote Electronics Enclosure - Optional Flameproof: Class I, Division 1, Groups B, C, D Temperature Codes: T2D (Ta=75° C), T3 (Ta=65° C) Class I, Zone 1, AEx D IIB + H2 Enclosure Typical 4/IP55

Communication

  • Modbus RTU Communication-Alarms, Flow Status & Diagnostics
  • RS-485 Using Kayden's Remote Control & Monitoring Software (RCMS) or other Modbus Compatible Software

Cover - for Local Enclosure

  • Blind Cover - Flameproof
  • Glass Lens Cover - Flameproof

Electronics - Standard Features

  • *Configurable only from the Kayden RCM Software
  • 100% Digital Design. Electronics are Environmentally Sealed
  • 4-Adjustable Independent Switch Point Timers
  • Automated Redundant Self-Test Diagnostics
  • Display Panel Lock-Out*
  • Easy Setup: No Mechanical Jumpers or Trim Pots
  • Incrementally Adjustable Heater Power (5% Steps) with Automatic Over-Range Protection
  • Intelligent User Interface with Push-Button & Constant Display of Operation (Flow/Level)
  • Start-Up Bypass Timer
  • Temperature Compensation
  • Temperature Mode*: Process Temperature may be Displayed/Graphed in Flow/Level Mode

Enclosure Finish

  • Powder-Coated Polyester TGIC

Enclosure, Body Details

  • 1" NPT Female Conduit Connection
  • Buna-N O-Ring on Cover

Enclosure, Body Material

  • Copper-Free Aluminum (Does Not Exceed 0.4% Copper)

Environmental Protection

  • IP55
  • IP67
  • NEMA 4X

Input Power

  • 12-24 Vdc, 120-240 Vac, 50/60 Hz
  • Power Consumption: Maximum
  • Universal Power Input

Insertion 'U' Length

  • 1.2" (3 cm)
  • 12" (30 cm)
  • 18" (46 cm)
  • 2" (5 cm)
  • 3" (7.5 cm)
  • 4" (10.1 cm)
  • 6" (15.2 cm)
  • 9" (23 cm)
  • Custom 'U' Lengths 2.5" to 120" (6.4 cm to 305 cm) Available in 1/2" (1.0 cm) Increments.

NPT Connection

  • 1" NPT Male
  • 1/2" NPT Male
  • 1-1/2" NPT Male
  • 1-1/4" NPT Male
  • 2" NPT Male
  • 3/4" NPT Male

Operating Pressure - Sensor

  • CRN - All Process Connections - 316/316 Stainless Steel - Canadian Registration Number 0F20731.5C: Maximum
  • CRN - All Process Connections - Hastelloy® C-276 - Canadian Registration Number 0F20731.5C: Maximum
  • CRN - All Process Connections - Titanium - Canadian Registration Number 0F20731.5C: Maximum
  • CSA - All Process Connections: Maximum

Operating Temperature - Electronics

  • -55° to 65° C (-67° to 149° F)

Operating Temperature - Sensor

  • -45° to 200° C (-50° to 392° F)

Output

  • 4-20 mA (Thermal Signal, Scaleable) 120 Vac, Maximum

Principle of Operation

  • Thermal, Calorimetric

Remote Control & Monitoring Software

  • Adjust Relay Modes
  • Adjust Set Points 1 & 2
  • Load or Save User Settings
  • Lock or Unlock the Display Panel to Protect the Settings
  • Many Other Adjustable Features
  • Modify Heater Settings to Optimize Response Time
  • Monitor Operating Status
  • Operated Via a Personal Computer Locally or from a Remote Control Center
  • Visualize the Actions of the Process Via the Graphical Display

Remote Electronics Enclosure & Cover

  • Blind Cover - Flameproof
  • Glass Lens Cover - Flameproof

Response Time

  • Approximately 0.5 to 30 seconds

Sensor, Probe Style

  • Insertion

Storage Temperature

  • Product Should Be Stored in a Clean & Dry Environment between -30° & 60° C (-22° & 140° F)

Switch Point Range

  • Gases:0.25 to 200 sfps, 0.08 to 60 SMPS - Standard Condition 21° C (70° F) at 14.7 psi (1 Atm)
  • Hydrocarbon-based Liquids: 0.01 to 5.0 FPS, 0.003 to 1.50 MPS
  • Water-based Liquids: 0.01 to 3.00 FPS, 0.003 to 0.9 MPS

Switch Type

  • SPDT Sealed Relay Contacts Rated @ 4 Amps Resistive 230 Vac or 30 Vdc, Maximum

Wetted Materials

  • 316, 316L Stainless Steel
  • All-Welded Construction
  • Hastelloy® C-276
  • Titanium

Documentation

Outline Drawings

Similar SKUs

810-RAC-I0170-C1G0A3E
810-RAC-I0170-C1G0A3E
Kayden CLASSIC® 810 Flow, Level, Interface Switch & Transmitter
Kayden Instruments
$568.87 USD
25+ In Stock
KAYDEN® CLASSIC® Thermal Flow, Level, Interface Switch, Modbus, 1/2" NPT, Custom 17"U Length, 316 SST Sensor, Local Electronics Enclosure with Glass Lens Cover, Approvals: CRN/CSA
810-RAD-I0090-C1B0A1E
810-RAD-I0090-C1B0A1E
Kayden CLASSIC® 810 Flow, Level, Interface Switch & Transmitter
Kayden Instruments
$105.14 USD
13 In Stock
KAYDEN® CLASSIC® Thermal Flow, Level, Interface Switch, Modbus, 3/4" NPT, 9"U Length, 316 SST Sensor, Local Electronics Enclosure with Blind Cover, Approvals: CSA
810-RAD-I0180-C1B0A1E
810-RAD-I0180-C1B0A1E
Kayden CLASSIC® 810 Flow, Level, Interface Switch & Transmitter
Kayden Instruments
$710.89 USD
8 In Stock
KAYDEN® CLASSIC® Thermal Flow, Level, Interface Switch, Modbus, 3/4" NPT, 18"U Length, 316 SST Sensor, Local Electronics Enclosure with Blind Cover, Approvals: CSA
810-RAC-0090-C1B0A1E
810-RAC-0090-C1B0A1E
Kayden CLASSIC® 810 Flow, Level, Interface Switch & Transmitter
Kayden Instruments
$265.21 USD
20 In Stock
KAYDEN® CLASSIC® Thermal Flow, Level, Interface Switch, Modbus, 1/2" NPT, 9"U Length, 316 SST Sensor, Local Electronics Enclosure with Blind Cover, Approvals: CSA
810-RAC-I0055-C1B0A3E
810-RAC-I0055-C1B0A3E
Kayden CLASSIC® 810 Flow, Level, Interface Switch & Transmitter
Kayden Instruments
$265.99 USD
25+ In Stock
KAYDEN® CLASSIC® Thermal Flow, Level, Interface Switch, Modbus, 1/2" NPT, 5.5"U Length, 316 SST Sensor, Local Electronics Enclosure with Blind Cover, Approvals: CRN/CSA
810-RAC-I0240-C1G0A1E
810-RAC-I0240-C1G0A1E
Kayden CLASSIC® 810 Flow, Level, Interface Switch & Transmitter
Kayden Instruments
$139.67 USD
25+ In Stock
KAYDEN® CLASSIC® Thermal Flow, Level, Interface Switch, Modbus, 1/2" NPT, 24"U Length, 316 SST Sensor, Local Electronics Enclosure with Glass Lens Cover, Approvals: CSA
810-RAC-M0860-C1B0A1E
810-RAC-M0860-C1B0A1E
Kayden CLASSIC® 810 Flow, Level, Interface Switch & Transmitter
Kayden Instruments
$714.81 USD
19 In Stock
KAYDEN® CLASSIC® Thermal Flow, Level, Interface Switch, Modbus, 1/2" NPT, 86 cm U Length, 316 SST Sensor, Local Electronics Enclosure with Blind Cover, Approvals: CSA
810-RAC-I0080-C1B0A1E
810-RAC-I0080-C1B0A1E
Kayden CLASSIC® 810 Flow, Level, Interface Switch & Transmitter
Kayden Instruments
$256.58 USD
23 In Stock
KAYDEN CLASSIC Thermal Flow,Level,Interface Switch, Modbus, 1/2" NPT, 8"U, SST, SST, Blind, CSA